Single Channel Signal Buffer Module
P/N: ASP-SBM-00
Signal quality problems caused by mismatched impedance can result in intermittent measurements and faulty diagnostics. This is most often observed when transporting a test program from one system to another, such as re-hosting it from an older generation system to a new one. The primary root cause of the impedance mismatch is the signal path from the UUT to the measurement instrument. The switching system technology often is responsible for much of the distortion.
The Signal Buffer Module (SBM) is designed to receive a signal from the UUT with high impedance output and accurately reproduced it at the output for delivery to the measurement instrument. The SBM output matches the 50 ohm input impedance of most measurement instruments and result in significantly reduced signal distortion. Hardware Specifications
Please contact us if you have questions or to request a quote. Example Waveforms
Fig 1. Channel 1 is a 20 MHZ clock at the UUT. Channel 3 is the signal at the Counter Timer
after it has passed through the Cross Point Matrix WITHOUT the SBM.
Fig 2. Channel 1 is a 20 MHZ clock at the UUT. Channel 3 is the signal at the Counter Timer after it has passed through the SBM and the Cross Point Matrix.
Fig 3. Channel 1 is the signal at the UUT. Channel 3 is the signal at the Counter Timer WITHOUT the SBM.
Fig 4.Channel 1 is the signal at the UUT. Channel 3 is the signal at the Counter Timer after it has passed through the SBM and the Cross Point Matrix.
|
||||||||